healthrian_ble_support_for_p30
Healthrian BLE support library for the Hativ P30 ECG device, built on
healthrian_ble_interface. It is a thin
device-configuration layer: the generic ModuleBle / ModuleMonitor transport
and streaming come from the interface; this package only supplies the P30's
UUIDs, command framing, and packet parsing.
Device profile
- Scan filter: advertised name starts with
Hativ. - Channels: 6 — the device streams 2 physical leads (L1, L2) and the remaining four limb leads (L3, aVR, aVL, aVF) are derived. No precordial (V1–V6) leads.
- Sampling: 250 Hz, 20 samples per BLE notification (24-bit signed, big-endian, scaled to mV).
- Command framing:
[0xBB, 0xCC, length, ...payload](no CRC). ECG start =0x10, stop =0x11. - Power-on: set mode
BB CC 02 0A 01, then aBB CC 01 FFkeep-alive every 2 s — the P30 sleeps without it. Declared as commands (setMode,powerOnMode,keepAlive,keepAliveInterval) and driven byModuleBle.powerOn(), whichregisterCharacteristicsruns on every connect. Consumers do not have to write these packets themselves.
| Characteristic | UUID |
|---|---|
| CONTROL_POINT | 00002CF0-0000-1000-8000-00805F9B34FB |
| MEASUREMENT_ECG | 00002C07-0000-1000-8000-00805F9B34FB |
| LEAD_STATE_ECG | 00002C00-0000-1000-8000-00805F9B34FB |
| BATTERY_LEVEL | 00002A19-0000-1000-8000-00805F9B34FB |
| FIRMWARE_VERSION | 00002A26-0000-1000-8000-00805F9B34FB |
| SERIAL_NUMBER | 00002A25-0000-1000-8000-00805F9B34FB |
Usage
import 'package:healthrian_ble_support_for_p30/healthrian_ble_support_for_p30.dart';
final manager = BLEManager(); // singleton
// Runs the P30 power-on sequence (set mode + 2 s keep-alive) on its own.
await manager.connectAndRegister(targetDevice: deviceId);
await manager.initializeStreams();
await manager.startMeasurement();
// 6-channel converted samples [channel][sample]
manager.ble.streams.ecg.meta.value?.listen((channels) {
// channels[0]=L1, [1]=L2, [2]=L3, [3]=aVR, [4]=aVL, [5]=aVF
});
manager.ble.streams.ecg.meta.leadState?.listen((state) { /* electrode contact */ });
await manager.stopMeasurement();
await manager.disconnectDevice();